Measurement functions |
Number of channels: 2, for temperature measurement with thermocouple (voltage measurement not available) |
Input connectors |
Thermocouple input: plug-in connector, Recommended wire diameter: single-wire, 0.14 to 1.5mm2, braided wire 0.14 to 1.0 mm2 (conductor wire diameter min. 0.18mm), AWG 26 to 16
Input impedance: min. 5 MΩ (with line fault detection ON/OFF), Max. rated voltage to earth: 300V AC, DC (with
input isolated from the unit, the maximum voltage that can be applied
between input channel and chassis and between input channels without
damage) |
Temperature measurement range Note: Upper and lower limit values depend on the thermocouple
|
10°C/div (-100°C to 200°C), 50°C/div (-200°C to 1000°C), 100°C/div (-200°C to 2000°C), 3ranges, full scale: 20div,
Measurement resolution: 1/1000 of measurement range (using 16-bit A/D conversion and when installed in the 8847) |
Thermocouple range (JIS C 1602-1995) (ASTM E-988-96) |
K: -200 to 1350°C, J: -200 to 1100°C, E: -200 to 800°C, T: -200 to 400°C, N: -200 to 1300°C, R: 0 to 1700°C, S: 0 to 1700°C,
B: 400 to 1800°C, W (WRe5-26): 0 to 2000°C,
Reference junction compensation: internal/ external (switchable),
Line fault detection ON/OFF possible |
Data refresh rate |
3methods, Fast: 1.2ms (digital filter OFF), Normal: 100ms (digital filter 50/60Hz), Slow: 500ms (digital filter 10Hz) |
Measurement accuracy |
Thermocouple K, J, E, T, N: ±0.1% of full scale ±1°C (±0.1% of full scale ±2°C at -200°C to 0°C),
Thermocouple R, S, W: ±0.1% of full scale ±3.5°C (at 0°C to 400°C or less), ±0.1% of full scale ±3°C (at 400°C or more)
Thermocouple B: ±0.1% of full scale ±3°C (at 400°C or more),
Reference junction compensation accuracy: ±1.5°C (added to measurement accuracy with internal reference junction compensation) |
Dimensions, mass |
106mm (4.17in) W × 19.8mm (0.78in) H × 204.5mm (8.05in) D, 240g (8.5oz) |
Supplied accessories |
Ferrite clamp × 2 |
|
|
 |
 |
 |
 |
 |
Features
Measure temperature graphs
- 16-bit A/D high resolution measurement
- Isolated input, up to 300 V AC/DC from the earth
|
|
 |
 |
 |
 |

|