■ Basic specifications (Accuracy guaranteed for 1 year)Allowable input voltage | Up to 5 V |
---|
Measured information | Impedance, voltage, temperature |
---|
Impedance measurement | Parameters: R, X, Z, θ, Frequency: 0.1 Hz to 1050 Hz, Measurement ranges: 3.0000 mΩ, 10.0000 mΩ, 100.000 mΩ Testing current: 3 mΩ range: 1.5 Arms, 10 mΩ range: 500 mArms, 100 mΩ range: 50 mArms |
---|
Voltage measurement | Measurement range: 5.00000 V (single range), Measurement time: 0.1 s (Fast) to 1.0 s (Slow) |
---|
Temperature measurement | Range: -10.0 °C to 60.0 °C, Measurement time: 2.3 s |
---|
Basic accuracy | Z: ±0.4% rdg. ±(8 to 60) dgt. , V: ±0.0035% rdg. ±5 dgt. , Temp.: ±0.5 °C (at 10.0 to 40.0 °C) |
---|
Functions | Comparator, self-calibration, sample delay, average, contact check, measurement current error, and other |
---|
Interfaces | RS-232C/USB (virtual COM port) * Cannot be used simultaneously EXT. I/O (NPN/PNP can be switched) |
---|
Power supply | 100 to 240 V AC, 50/60 Hz, 80 VA max |
---|
Dimensions and mass | 330 mm (12.99 in) W × 80 mm (3.15 in) H × 293 mm (11.54 in) D, 3.7 kg (130.5 oz) |
---|
Accessories | Power cord ×1, Instruction manual ×1, Zero-adjustment board ×1, USB cable (A-B type) ×1, CD-R (communication instruction manual, PC application software, USB driver) ×1 |
---|
|
|
|
|
|
|
|