[Withstanding
test] |
|
|
|
Testing
voltage |
0.2V AC to 5.00kV AC |
Voltage
setting method |
Digital setting, Setting
resolution: 0.01kV |
Waveform/
Frequency |
Sine wave (Distortion ratio 5%
or less at no load), 50/60Hz selectable |
Current
measurement range |
0.01mA to 20.0mA, True RMS
rectified (digital display) |
Measurement
range |
10mA (0.01mA resolution), 100mA
(0.1mA resolution) |
Voltage
meter |
Accuracy: ±1.5% rdg. (1000 V or
more), ±15V (less than 1000V), True RMS rectified |
Judgment
function |
Window comparator method
(Digital setting) |
|
[Insulation
test] |
|
|
|
Testing
voltage |
500V DC, 1000V DC |
Rated
testing current |
1 to 1.2mA, Shorted current: 4
to 5mA (at 500V), 2 to 3mA (at 1000V) |
Measurement
range, Accuracy |
0.5MΩ to 999MΩ (at 500V), and
1MΩ to 999MΩ (at 1000V) : ±4% rdg., 1000MΩ to 2000MΩ : ±8% rdg. |
Judgment
function |
Window comparator method
(Digital setting) |
|
[Timer] |
|
|
|
Measurement
range, Accuracy |
0.3 to 999s, and 0.3 to 99.9s :
±50ms, 100 to 999s : 0.5s |
Delay/Ramp
timer |
0.1 to 99.9s: ±50ms, 0.1s
resolution |
Functions |
Saving eight testing conditions,
hold, buzzer, contact check |
Moniter
function |
Output voltage, detected
current, insulation resistance, Refresh rate: 4times/s |
Power
consumption |
100 to 240V AC, (50/60Hz), 200VA
max. |
Dimensions,
mass |
320mm (12.60in) W × 155mm
(6.10in) H × 395mm (15.55in) D, 15kg (529.1oz) |
Supplied
accessories |
H.V. TEST LEAD 9615 (Red, black)