■ Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)Regarding DMM Accuracy Due to the many ranges and functions available in a DMM, only the basic accuracy is indicated for reference. Please refer to the individual catalogs for detailed accuracy information. DC Voltage (Basic accuracy) | 600.0 mV to 1000 V 5 ranges, Basic accuracy: ±0.3 % rdg. ±3 dgt. |
---|
AC Voltage 40 to 500 Hz | 6.000 V to 1000 V, 4 ranges, Frequency characteristics: 40 Hz to 1 kHz Basic accuracy: ±0.9 % rdg. ±3 dgt. (True RMS, crest factor 3) |
---|
AUTO AC/DCV | Yes |
---|
Resistance | 600.0 Ω to 60.00 MΩ, 6 ranges, Basic accuracy: ±0.7 % rdg. ±3 dgt. |
---|
DC Current Basic accuracy | 60.00 mA to 10.00 A, 4 ranges, ±0.9 % rdg. ±3 dgt. |
---|
From 4 to 20 mA% conversion display | Yes |
---|
AC Current (direct input) 40 to 500 Hz Basic accuracy | 600.0 mA to 10.00 A, 3 ranges, ±1.4 % rdg. ±3 dgt. (True RMS, crest factor 3, 40 Hz to 1kHz) |
---|
AC Current (use with Clamp on probes) 40 to 1k Hz | 10.00 A to 1000 A, 7 ranges, Add the clamp on probe accuracy to basic accuracy: ±0.9 % rdg. ±3 dgt. (True RMS, crest factor 3) |
---|
Voltage detection | Hi: AC40 V to 600 V, Lo: AC80 V to 600 V |
---|
C (capacitance) | 1.000 μF to 10.00 mF, 5 ranges, Basic accuracy: ±1.9 % rdg. ±5 dgt. |
---|
Frequency | 99.99 Hz (5 Hz or more) to 99.99 kHz, 4 ranges (limited by the minimum detectable voltage and current), Basic accuracy: ±0.1 % rdg. ±1 dgt. |
---|
Continuity check | Continuity threshold [ON]: 25 Ω or less (Indicate buzzer sound, red LED), [OFF]: 245 Ω or more, Response time: 0.5 ms or more |
---|
Diode test | Open terminal voltage: 5.0 V or less, Testing current 0.5 mA or less, Threshold of forward voltage: 0.15 V to 1.5 V |
---|
Other functions | Filter function, Display value hold, Auto hold, Max/Min/Average value display, Relative display, Auto-power save, USB communication (option) |
---|
Display | Main and Sub displays: 4-digits LCD, max. 6000 digits, bar graph |
---|
Display refresh rates | 5 times/s (Capacitance measurement: 0.05 to 5 times/s, depending on measured value, Frequency: 1 to 2 time/s, Temperature: 1 time/s ) |
---|
Power supply | LR03 alkaline batteries ×4, Continuous use: 130 hours (backlight OFF) |
---|
Dimensions and mass | 84 mm (3.31 in)W × 174 mm (6.85 in)H× 52 mm (2.05 in)D, 390 g (13.8 oz) (including batteries and holster) |
---|
Accessories | Test lead L9207-10 ×1, Holster ×1, Instruction manual ×1, LR03 alkaline battery ×4 |
---|
|
|
|
|
|
|
|
|
|
|
|
|
|
Features
The Hioki DT4200 Digital Multimeter Series consist of a complete lineup of 9 models ranging from professional to industrial to pocket meters, delivering a super fast response rate and safety features that take electrical testing to a higher level. The DT4256 is a digital multimeter designed to deliver maximum functionality for use in a wide range of settings, including instrumentation testing with a 4-20 mA% conversion function.
• 11 functions including 4-20 mA% conversion • ±0.3% DC V basic accuracy, wide 40 Hz to 1 kHz AC V frequency characteristics • Low-pass filter (100 Hz/ 500 Hz) cuts high harmonics(when measuring inverter fundamental waveforms) • Connect to PC with USB communication function (optional) • Broad -10 (14°F) to 50°C (122°F) operating temperature range
|
|
|
|
|
|
|