DC GENERAL | MEASUREMENT TIME CHOICES | 0.01 ~ 10PLC1,0.01PLC/step |
| AVERAGE READINGS | 1~10 |
| TYPICAL READING TIME2,3 | 31ms |
DC VOLTAGE OUTPUT (23℃±5℃) | OUTPUT VOLTAGE | 0~15V |
| OUTPUT ACCURACY | ± (0.05%+10mV) |
| PROGRAMMING RESOLUTION | 2.5mV |
| READBACK ACCURACY3 | ± (0.05%+3mV) |
| READBACK RESOLUTION | 1mV |
| OUTPUT VOLTAGE RISING TIME | 0.15ms (10% ~ 90%) |
| OUTPUT VOLTAGE FALLING TIME | 0.65ms (90% ~ 10%) |
| LOAD REGULATION | 0.01%+2mV |
| LINE REGULATION | 0.5mV |
| STABILITY4 | 0.01%+0.5mV |
| RECOVERY TIME(1000%LOAD CHANGE) | <40us (<100mV) <80us (<20mV) |
| RIPPLE AND NOISE5 | 1mV rms(0~1MHz) 8mVpp(20Hz~ 20MHz) |
DC CURRENT (23℃±5℃) | OUTPUT CURRENT | 0 ~ 5A (0 ~ 9V) 0 ~ 3A (9 ~ 15V) |
| SOURCE COMPLIANCE ACCURACY | ±(0.16%+5mA) |
| PROGRAMMED SOURCE RESOLUTION | 1.25mA |
| READBACK ACCURACY3 | 5A range: ±(0.2%+400uA) 5mA range: ±(0.2%+1uA) |
| READBACK RESOLUTION | 5A range: 100uA 5mA range: 0.1uA |
| CURRENT SINK CAPACITY | 0 ~ 5V: 2A 5 ~ 15V:(2A derate 0.1A)/V |
| LOAD REGULATION | 0.01%+1mA |
| LINE REGULATION | 0.5mA |
| STABILITY4 | 0.01%+50uA |
DVM | INPUT VOLTAGE RANGE | 0 ~ 20VDC |
| INPUT IMPEDANCE | 10Ω |
| MAXIMUM INPUT VOLTAGE | -3V, +22V |
| READING ACCURACY3 | ± (0.05%+3mV) |
| READING RESOLUTION | 1mV |
PULSE CURRENT MEASUREMENT | TRIGGER LEVEL | 5mA ~ 5A, 5mA/step |
| HIGH TIME/LOW TIME/AVERAGE TIME | 33.3us to 833ms, 33.3us/step |
| TRIGGER DELAY | 0 ~ 100ms,10us/steps |
| AVERAGE READINGS | 1 ~ 100 |
| LONG INTEGRATION PULSE TIMEOUT | 1S ~ 63S |
| LONG INTEGRATION MEASUREMENT TIME | 850ms(60Hz)/840ms(50Hz) ~ 60s,or AUTO time 16.7ms/steps(60Hz), 20ms/steps(50Hz) |
| LONG INTEGRATION TRIGGER MODE | Rising, Falling, Neither |
OVP | OVP RANGE | OFF,ON (0 ~ 15.2V) |
| RESOLUTION | 10mV |
| ACCURACY | 50mV |
Others | PROGRAMMING | IEEE-488.2(SCPI) |
| USER_DEFINABLE POWER_UP STATES | 5 sets |
| REAR PANEL CONNECTOR | 8Pin:output*4, sense*2, DVM*2 |
| TEMPERATURE COEFFICIENT | 0.1* specification/ ˚C |
| POWER CONSUMPTION | 150VA |
| REMOTE/LOCATION CONNECTOR | USB/GPIB/LAN |
| RELAY CONTROL CONNECTOR | 150mA/15V 5Voutput, 100mA |
Insulation | Chassis and Terminal | 20MΩ or above (DC 500V) |
| Chassis and AC cord | 30MΩ or above (DC 500V) |
Operation Environment | | Indoor use, Altitude: ≤ 2000m Ambient temperature: 0 ~ 40°C Relative humidity: ≤ 80% Installation category: II, Pollution degree: 2 |
STORAGE Environment | | TEMPERATURE: -20˚C ~ 70˚C HUMIDITY: < 80% |
INPUT POWER | | 90-264VAC, 50/60Hz 6 |
Accessories | | CD 8cmUser manual x1, Quick Start manual x1 Test lead GTL-117 x 1, GTL-203A x 1, GTL-204A x 1 |
Dimensions | | 222 (W) x 86 (H) x 363 (D) mm |
Weight | | Approx. 4.2kg |
| | |
Remarks | | - PLC=PowerLineCycle, 1PLC = 16.7ms for 60Hz operation, 20ms for 50Hz operation;
- Display OFF, Speed includes measirement and binary data transfer out of GPIB;
- PLC=1;
- STABILITY:Following 15 minute warm-up, the change in output over 8 hours under ambient temperature, constant load, and line operating conditions;
- The ground ring of the probe is pressed directly against the output ground of the power supply and the tip is in contact with the output voltage pin.
- Auto detected at power-up;
|