■ Basic specifications (Accuracy guaranteed for 1 year)Measurement methods | Measurement of voltage drop across body simulated resistance points, Calculation and display of current values, True rms measurement, Measurement unit floats relative to instrument ground. |
---|
Measurement modes | Leak current measurement, voltage measurement, safety conductor current measurement |
---|
Standard compliance | Electrical Appliances and Materials Safety Act Measurement of touch current and protective conductor current: IEC 60990 (1999) Electrical equipment for measurement, control, and laboratory use: IEC 61010-1 (2001) Information technology equipment: IEC 60950-1 (2005) Household and similar appliances: IEC 60335-1 (2001) + A1:2004 + A2:2006 Audio, video and similar electronic apparatus: IEC 60065 (2001) + A1:2005 Personnel Protection Systems for EV: UL-2231-1 (2002), UL-2231-2 (2002) UL: UL-1492 (1996) Electrical equipment for measurement, control, and laboratory use; current measurement circuits in damp conditions: IEC 61010-1 (2001) |
---|
Leak current measurement | Ground leak current, 3 types of contact current, free current measurement, 3 types of enclosure leak current |
---|
Measurement current | DC, AC (true rms, 15 Hz to 1 MHz), AC+DC (true rms, 15 Hz to 1 MHz), AC peak (15 Hz to 1 MHz) |
---|
Measurement ranges | DC / AC / AC+DC mode:50 μA/500 μA/5 mA/50 mA AC peak mode:500 μA/1 mA/10 mA/75 mA |
---|
Measurement accuracy (current measurement) | DC measurement: ±2.0% rdg. ±6 dgt. (typ.) AC / AC+DC measurement: ±2.0% rdg. ±6 dgt. (15 Hz to 100 kHz, typ.) AC peak measurement: ±2.0% rdg. ±6 dgt. (15 Hz to 10 kHz, typ.) |
---|
Interfaces | External I/O, USB 1.1 (communications), RS-232C |
---|
Functionality | Automatic test, data storage for 100 target devices, clock, data backup, printed output (optional), etc. |
---|
Power supply | 100/120/220/240 V AC (specify at time of order), 50/60 Hz, 30 VA rated power |
---|
Target device power supply input | 100 to 240 V AC, 50/60 Hz, Rated current input from terminal block: 20 A |
---|
Target device power supply output | Output from terminal block: 20 A, Output from outlet: 15 A |
---|
Dimensions and mass | 320 mm (12.60 in) W × 110 mm (4.33 in) H × 253 mm (9.96 in) D, 4.5 kg (158.7 oz) |
---|
Accessories | Test Lead L2200 (Red ×1, Black ×1) ×1 set, Enclosure Probe 9195 ×1, Power cord ×3, Spare fuse for measurement line ×1, Instruction manual ×1, Users guide ×1, CD-ROM ×1
|
---|
|
|
 |
 |
 |
 |

|